Abstract
Ethylene (C2H4) was detected using quartz-enhanced photoacoustic spectroscopy (QEPAS) at 10.5 µm with a continuous wave, distributed-feedback quantum cascade laser as the light source. The QEPAS sensor was operated at low pressures (≤200 torr) to eliminate the cross-talk spectral interference between C2H4 and CO2, a major interfering species in practical applications. The sensor was calibrated to show a good linear response to C2H4 concentration and the Allan deviation analysis demonstrated a minimum detection limit of 8 ppb at an integration time of 90 s. Although no spectral overlap between C2H4 and CO2 was confirmed at the pressure ≤200 torr by the direct absorption measurement using a 28-m multipass cell, we observed the apparent influence of the CO2 addition to the C2H4/N2 mixture on the photoacoustic signal of C2H4. An energy transfer model involving the vibration–vibration (VV) and vibration–translation (VT) transitions in the C2H4–CO2–N2 system was constructed to interpret the experimental data. Additionally, the vibrational relaxation times of C2H4 were obtained based on the QEPAS technique and the energy transfer model, which were in good agreement with the previous studies.
© 2017 The Author(s)
PDF Article
More Like This
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator